Maintenance
& Troubleshooting
.Replacement procedures
Pentra
80 - User Manual - RAB108EA
7–5
▼
Sampling probe
Sampling probe must be decontaminated as follows:
1- Prepare a solution of Sodium Hypochlorite to 100ml/l*.
2- Fill a 5ml tube with this solution.
3- Run 5 analysis on bleach, using the «Stat» mode as described in (See Section 2,
Please also refer to the W.H.O (World Health Organization) guidelines: «Laboratory
Biosafety Manual, 2nd edition», for further information
2.
.
Replacement procedures
This section describes how to replace:
◆
a low level reagent (see
◆
an optical bench lamp (see
2.2. Optical bench lamp replacement,
◆
the sampling probe (see
2.3. Sampling probe replacement,
2.1. Reagent replacement
2.1.1. Reagent locations and connections
When installing the Diluent on the system, it is most important to maintain the depth
of the container no lower than 80 cm (31.5 in.) below the instrument. If the depth
limit is exceeded, erroneous results will occur.
Diluent and Waste tubing lengths are critical upon installation. Follow the recom-
mended lengths for best results:
DILUENT input tubing: Cristal 3x6/ 2 meters (80 in.) maximum.
WASTE output tubing: Cristal 4x6/ 2 meters (80 in.) maximum.
Bottles and container locations (see
1- ABX Lyse
2- ABX Basolyse II
3- ABX Eosinofix
4- ABX Cleaner
5- ABX Diluent
6- Waste container
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...