5.3.4. LMNE matrix flags
▼
Suspiscion
◆
When WBC populations are detected in abnormal quantities in one or more areas of the
LMNE matrix, a «
!
» suspect flag will be displayed next to the parameter(s) in question. If
one result appears with one or several parameters that display a «
!
»suspect flag, the result
should be further investigated. (Pathology suspicion, clotted sample, plasma cells, etc…).
▼
NO flag
This alarm appears in the «Analyzer Alarms» area on the screen and the printout.
Fig. 4–21 No flag
Standard value for NO: %100#120
8.3. Alarms & Curve thresholds
Meaning: Background
NO
ise.
This flag occurs when the number of parti-
cles counted in the background noise area
is higher than the limit set up in
NO#
or
when the number of counted particles ver-
sus the total number of WBC, is above the
NO%
limit.
Suspected abnormalities:
• Platelet aggregates,
• Large number of platelets,
• Erythrocyte membrane resistant to lysis
(stroma),
• NRBCs,
• Pollution.
NoE
NoN
Channel 127
NE
Absorbance
EOS
Left
Neu
Neutro
LN
RN
Right
Neu
RMN
Right
Mono
Mono
Aty.
Lym
Lympho
Left
Lym
NoL
LL
AL
LMD RM
Resistivity
NL
NO
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...