Procedure 10: Differential (LMNE)
Repeatability (If Differential results are non repeatable)
a – Perform a concentrated cleaning (see
b – If this does not correct the Differential count, call your ABX Diagnostics representative
department.
6.3. Flags
▼
Procedure 11: Default analysis
WBC:
a – Perform a concentrated cleaning (see
b – Re-run the specimen.
c – Check the operation of liquid valve <23> and <14> (Opening and closing during cycle).
If defective replace the valve.
d – If this does not correct the WBC results, call your ABX Diagnostics representative de-
partment.
RBC, PLT:
a – Perform a concentrated cleaning (see
b – Re-run the specimen.
c – Check the operation of liquid valve <14> (Opening and closing during cycle). If defec-
tive replace the valve.
d – If this does not correct the RBC\PLT results, call your ABX Diagnostics representative
department.
HGB:
a – Check the HGB LED is illuminated when system is on:
Fig. 7–45 HGB LED
b – Perform a concentrated cleaning (see
c – Re-run the specimen.
d – Check the operation of liquid valve <14> (Opening and closing during cycle). If defec-
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...