Set
tings
Rules
Pentra
80 - User Manual - RAB108EA
5–13
4. Rules
4.1. Accessing the «Rules» screen
From the «Settings» window, select the «Rules» key
Fig. 5–10 Accessing the Rules screen
The «Rules» screen includes three tabs which will enable the operator to configure the fol-
lowing conditions:
◆
Rerun conditions (see
◆
Print conditions (see
◆
Transmit conditions (see
4.2. Rerun conditions
In this tab, the user has the option of establishing the criteria for sample rerun conditions.
The re-sampling of sample tubes will be based on these criteria.
Re-sampling will be required if:
a – hematological results trigger alarm conditions (L1, LL1, MB...). These criteria are de-
fined for all types of blood (see
b – hematological results are beyond their parameter limits, which are defined by the user,
or samples that are rejected. These criteria are specific to each sample Type
(see
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...