Maintenance
& Troubleshooting
Super User menu description
Pentra
80 - User Manual - RAB108EA
7–25
Open the instrument right and left front doors and remove the right-hand and left-hand side
panels (see
3. Instrument panels & cover Removals,
Once the side panels have been removed, power on the instrument. From the Main screen,
select the «Super User Menu» and then
/Mechanical /Check Valves.
Closely observe the valve operations; movements have to be straight and regular.
Fig. 7–30 Super User\Mechanical\Check Valves
1-
Valves 1 to 12
: Verify the correct operation of these valves
2-
Valves 13 to 19
: Verify the correct operation of these valves
3-
Valves 20 to 23
: Verify the correct operation of these valves
4-
Valves 24 to 30
: Verify the correct operation of these valves
5-
Valves 31 to 35
: Verify the correct operation of these valves
5.1.4. Check Sensors
Verify the activity of all of the mechanical sensors (mainly switches) on the instrument. If
all switches are working properly, they will be indicated in «Green».
From the Main screen, select the «Super User Menu» and then
/Mechanical /Check Sensors.
1
2
3
4
5
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...