Pentra
80
4–12
Pentra
80 - User Manual - RAB108EA
1.7.1. Sample identification on both barcode and Rack/Position
If a sample tube is identified by a barcode, although the setting is «
Rack / Position
, page 5-7, the instrument checks that the worklist Sample
ID corresponds to the tube barcode identification, and that the rack/position of the tube cor-
responds to the one entered in the Worklist. If not, the instrument generates an exception.
Example:
The tube, identified «123» position 1/1, does not correspond to the order position 1/1 which
is identified «124».
◆
This tube is analyzed in the instrument default test.
◆
The result and the order must be manually matched.
1.7.2. Identification with barcode, without order
However
PENTRA 80
does not generate an exception in this case, provided that the rack/po-
sition of this tube, has not been reserved for an order in the worklist.
Example:
◆
This tube is analyzed in the instrument default test.
◆
If the setting is «
Manual match on Exception = ON
, page 5-7, the Sample ID provided with the result is the barcode.
RACK
Rack: 1
Position: 1
Barcode: 123
WORKLIST
Rack:1
Position: 1
Sample ID: 124
RACK
Rack: 1
Position: 1
Barcode: 123
WORKLIST
Rack:
Position:
Sample ID:
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...