Maintenance
& Troubleshooting
Troubleshooting
Pentra
80 - User Manual - RAB108EA
7–37
Fig. 7–43 Troubleshooting guide
6.1. Instrument operation mode
▼
Procedure 1: Printer
See printer’s user manual to connect, to switch on/off or to feed paper.
▼
Procedure 2: Reagents
Bottle replacement (see
2.1.2. Integrated reagents & Diluent container replacement,
Waste container (see
2.1.3. Waste container replacement,
▼
Procedure 3: Instrument startup
Startup failed:
a – Check all reagent’s expiration dates: Replace bottle if necessary.
b – Re-run a Startup cycle.
c – Perform a concentrated cleaning (see
Temperature not reached:
a – Wait for five minutes to reach the operating temperature
b – If temperature is not reached call your ABX representative service department
Calibration verification out of acceptable limits:
a – Clean the system (see
page 7-33) and re-run the control.
b – Run a new vial.
c – Calibrate the instrument Section 3,
▼
Procedure 4: Sampling probe
Sampling probe:
a – Check probe motion (see
b – Open the righthand side panel to access to the chambers Section 7,
c – Run an analysis cycle on blood.
d – Control the specimen aspiration (Blood delivering in the chambers).
e – Check the probe is not bent.
▼
Procedure 5: Dilution
Carriage motion:
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...