Pentra
80
4–44
Pentra
80 - User Manual - RAB108EA
▼
RBC messages
Tab. 4–7: RBC pathology messages
«H»: extreme high limit
«L»: extreme low limit
Message
Condition
Anemia
HGB < HGB L
Anisocytosis
RDW > RDW H
Microcyte
MIC
Mic
% MIC > 10%
Mic+
% MIC > 15 %
Macrocyte
on Mac Flag
Hypochromia
MCHC < MCHC L
Cold Agglutinin
MCHC > MCHC H and WBC < 91.3x10
3
/mm
3
Microcytosis
MCV < MCV L
Macrocytosis
MCV > MCV H
Erythrocytosis
RBC > RBC H
Interpretation Not possible
RBC < 0,01x10
6
/mm
3
or RBC reject ( or RBC>0.03 during Startup)