Set
tings
Rules
Pentra
80 - User Manual - RAB108EA
5–15
4.3. Print conditions
From the «Rules» screen, select the «Print conditions» tab.
Fig. 5–13 Print conditions
Selection of printed results can have the following conditions:
◆
Manual: only when the operator requires a printed result from the «Results» screen.
◆
Unconditionnal: All results will be printed
◆
Selected criteria: the result must respond to one (or several) of the following conditions
to be printed:
1- one of the checked flags has been triggered
2- a rejected result occurs on one of the checked parameters
3- the results are out of panic or normal limits on one of the checked parameters.
To select or de-select criteria, refer to the «Rerun» tab. (see