Pentra
80
5–14
Pentra
80 - User Manual - RAB108EA
4.2.1. Rerun on alarms
4.2.2. Rerun by parameters
Fig. 5–11 Rerun on alarm
Refer to
page 4-29 for Hematologi-
cal alarms interpretation prior to selection in
this screen.
For modification of the Rerun conditions in
Alarms:
◆
Select the «Edit» key.
◆
Select the alarm box that you want to trig-
ger the re-sampling.
◆
Select the «OK» key to confirm your choice
of re-sampling conditions
Fig. 5–12 Rerun by parameters
From the «Rerun» screen,
scroll the «Types» list, and
select a type to configure.
Select the «EDIT» key.
Choose the limits (Panic,
Normal set by the user) or
rejected parameters that
will trigger a rerun.
Select or De-select the
boxes to trigger or not trig-
ger a rerun on parameter
limits.
If you want a rerun for any
parameter that is out of the
normal low limits, select
the «
l
» key and select any
of the parameters that you
would like to have a rerun
on low limits.
Now select the «OK» key to
confirm your selections
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...