Work
flow
Run results and associated Flags
Pentra
80 - User Manual - RAB108EA
4–31
5.3.3. Analysis reject
A reject flag (shown by *) occurs when two counts on a parameter differ more than the pre-
defined limits. It indicates that the parameter result is inconclusive and should be investi-
gated for (Manual) rerun status and/or instrument malfunction if consistent on every sample.
▼
RBC
If the RBC parameter is rejected, the MCV, MCH, MCHC, and RDW parameter results are re-
placed by «---» and/or have the «*» indicator.
▼
WBC
If the WBC parameter is rejected, the diff. parameter results are replaced by «---» and/or have
the «*» indicator .
▼
PLT
If the PLT parameter is rejected, the PCT, MPV, and PDW parameter results shall be replaced
by «---» and/or have the «*» indicator.
▼
Hemoglobin
◆
A «!» suspect flag is generated if two consecutive Hemoglobin blank measurements are dif-
ferent from the pre-defined value in «
Settings / Type Parametering / Alarms and Curves
Threshold Tab / Alarm Level Grid / HGB parameter absolute value
◆
A «!» suspect flag is generated if the three Hemoglobin measurements are different from
the pre-defined value in «
Settings, Type Parameterising, Alarms and Curves Threshold Tab,
Alarm Level Grid, HGB parameter percent value
◆
If 3 consecutive «
!
» suspect flags are noted on Hemoglobin during «Sample Analysis», the
Hemoglobin value will be rejected and replaced with «---».
◆
During instrument «STARTUP», the HGB Blank measurement value is pre-defined from fresh
Diluent. If the Blank value is not within the measured limits, the message «
Startup Failed:
HGB Blank
» will be displayed.
▼
LMNE
◆
A reject on the LMNE channel indicates a poor correlation between the resistive and the
optical measurements on the matrix. It is shown by a «*» on all the differential parameters
in % and #. The results are not reliable and specimen must be re-analyzed.
◆
NO flag: (see
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...