3. Measuring principles
3.1. Multi distribution sampling sytem (MDSS)
3.1.1. CBC mode
While in the CBC Mode,
30µl
of whole
blood is aspirated then delivered with
into the following chambers as
indicated:
One segment of sample for the Dilution
chamber which is used for the RBC/PLT
dilution and the Hemoglobin
measurement.
The other segment of sample is used for
the WBC/BASO count.
Fig. 6–10 Segment distribution in CBC mode
3.1.2. Diff Mode
While in the DIFF Mode,
53µl
of whole
blood is aspirated then delivered with
into the following chambers as
indicated:
One segment of sample for the Dilution
chamber which is used for the RBC/PLT
dilution and the Hemoglobin
measurement.
The second segment of sample is used
for the WBC/BASO count.
The last segment of sample is used for
the LMNE chamber from which the sam-
ple is taken into the flowcell for the
LMNE count.
Fig. 6–11 Segment distribution in DIFF mode
Diluent
Air
7µl discarded
10µl WBC/BASO
10µl RBC/PLT/HGB
3µl discarded
Diluent
Air
5µl discarded
10µl WBC/BASO
10µl RBC/PLT/HGB
3µl discarded
25µl Differential
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...