8.2. Pathological limits
Fig. 5–35 Pathological limits tab
A set of limits is available for each sample type. The operator, according to the laboratory
specifications, can modify these limits. The limits that have been entered into the 8 different
types (Standard, Man, Woman, Child1, Child2, Child3, Child4, and Child5) are Factory Default
values.
The «Standard Type» values are set values and cannot be modified by the user.
Man, Woman, Child1, Child2, Child3, Child4, and Child5 Factory Default values can be modi-
fied by the user.
◆
Results exceeding the «
Normal Ranges
» limits are marked with the following flags:
– «
h
» for results above the normal upper limit,
– «
l
» for results below the normal lower limit.
◆
Results exceeding the «
Panic Ranges
» limits are marked with the following flags:
– «
H
» for results above the extreme
upper limit,
– «
L
» for results below the extreme lower limit.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...