Maintenance
& Troubleshooting
.Replacement procedures
Pentra
80 - User Manual - RAB108EA
7–17
9- Press the «Validate» button then repeat this cycle a second time.
10- Run several manual cycles.
11- Remove the straws from the distilled water bottle then plunge them into an
empty bottle.
12- Run several «ALL» Unprime cycles (See
, page 16) to drain
the instrument.
2.5.2. Syringes and carriage park
Move the syringes and the carriage in a safe position:
1- Enter:
Service \ Super User Menu \ Others
and press the «Run park syringe
position» button (See
Fig. 7–20 Park screen
2- Before the transport of the instrument press the «Run maintenance carriage
position» button.
3- Block the carriage using the «Plastic rails» (See
Fig. 7–21 Plastic rails
4- Close the door and turn off the instrument.