80 - User Manual - RAB108EA
2.5. Instrument Rinse
Run this procedure before the transport of the instrument, after a demonstration or before a
long period without functioning.
2.5.1. Instrument rinse
1- Run an Instrument general cleaning (See
2- Remove straws from Reagent bottles and plunge them into an empty bottle.
3- Remove straw from Diluent cubitainer and plunge it into the empty bottle.
4- Enter:
Service \ Super User Menu \ Hydraulic \ Unprime Cycle
then click the
«All» button (See
Fig. 7–18 Unprime cycle screen
5- Press the «Validate» button then repeat this cycle a second time.
6- Dry the straws using absorbant paper.
7- Plunge the straws in a bottle full of distilled water.
8- Enter:
Service \ Super User Menu \ Hydraulic \ Prime Cycles
then click the «ALL
Fig. 7–19 Prime cycles screen