Pentra
80
5–38
Pentra
80 - User Manual - RAB108EA
8.3.2. 5DIFF Matrix thresholds
Each axis of the matrix (X and Y) is divided into 128 channels numbered from 0 to 127.
13 (Y) vertical indices and 13 (X) horizontal indices will allow the user to locate these chan-
nels by multiples of 10. The first index channel of the 5-Diff matrix is the origin (at the
bottom left corner). This is the «0» channel. The second channel will be 10, the third channel
will be 20, the fourth channel will be 30, and so on. The threshold adjustments are expressed
in channels.
Fig. 5–37 Matrix thresholds
There are only 3 basic reasons to adjust the matrix thresholds. (see
a – To improve the separation between different cell populations which may vary according
to the selection of anti-coagulant for blood sample collection or minor internal instru-
ment adjustments.
b – To modify the alarm areas of these cell populations for improvement in detection sen-
sitivity.
c – To modify one or more matrix areas in order to define a more precise measurement of
a specific cell population for research purposes.
NoE
NoN
LN
LL AL
LMD
RM
RN
LMN
NoL
LMU
MN
NL
RMN
NE
FLN
FMN
FNE
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...