Pentra
80
8–6
Pentra
80 - User Manual - RAB108EA
2.8. Reagent consumption (ml)
Tab. 8–3: Reagent consumption
STARTUP cycle estimated duration and consumptions are given for one blank cycle
control. It could be a maximum of three cycles.
2.9. Recommended Anticoagulant
The recommend
ed anticoagulant is K
3
EDTA, with the proper proportion of blood to
anticoagulant as specified by the tube manufacturer. K
2
EDTA is an acceptable
alternative.
Cycles
Estimated
duration (s)
Diluent
(ml)
Eosinofix
(ml)
Basolyse II
(ml)
Cleaner
(ml)
Lyse
(ml)
CBC/DIFF
0’45’’
27.4
1.0
2.0
1.0
0.45
CBC
0’45’’
24.4
-
2.0
1.0
0.45
Prime DILUENT
3’00’’
44
-
-
-
-
Prime
EOSINOFIX
1’34’’
1.6
23.7
-
-
-
Prime
BASOLYSE 2
1’25’’
1.7
-
23.7
1.0
-
Prime CLEANER
1’24’’
1.7
-
-
24.7
-
Prime LYSE
1’31’’
2.7
-
-
-
8.4
Prime ALL
7’13’’
50.7
24.0
24.0
25.0
8.4
STARTUP
(1 blank cycle)
2’28’’
55.2
2.0
3.0
2.0
0.95
SHUT DOWN
2’56’’
32.2
1.0
1.0
14.2
0.49
Rinse
CYTOMETER
1’14’’
5.0
-
-
-
-
AUTOCLEAN
1’33’’
27.6
1.0
1.0
1.0
0.50
MINICLEAN
0’21’’
10.3
1.0
2.0
1.0
0.33
CONCENTRATED
cleaning
4’00’’
29.6
1.0
1.0
1.0
0.50
BACKFLUSH
0’24’’
-
-
-
-
-
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...