Maintenance
& Troubleshooting
.Replacement procedures
Pentra
80 - User Manual - RAB108EA
7–11
Fig. 7–8
Reagent priming
2.1.3. Waste container replacement
Unscrew the waste container cap.
Replace waste container according to your laboratory’s protocol.
Close the empty container with the cap and dispose of waste liquids according to your lo-
cal\national organizations.
2.2. Optical bench lamp replacement
2.2.1. Lamp replacement
Wait for the lamp to cool down before handling it.
Power off the instrument.
Open the instrument left-hand and right-hand side panels, lift the top cover of the instrument
as well (see
3. Instrument panels & cover Removals,
Locate the Optical bench at the top left of the instrument
(