Page 4
FA2i-045A
4) Insert the 5-prong plug into the power receptacle on the lift mechanism as shown in Figure 5.
Figure 5 Insert the 5-Prong Plug
Notes:
Prior to installing the HFA table top onto the support tray:
•
Ensure that the correct fuses are installed for the power mechanism to be tested.
•
Plug the power cord into the power inlet on the lift mechanism and then plug the power cord into
any available power source (see label on the power lift mechanism for the proper power input
requirements).
WARNING –
Do not touch any exposed wire or terminal while the power cord is
plugged into the power source.
Notes:
•
Press the up and down button to test the lift mechanism to ensure that the 5-prong plug is
properly seated in the receptacle.
•
During testing, press the up button and allow the lift mechanism to reach its maximum extension.
Disconnect the power cord from the power inlet and observe that there is no slippage of the lift
mechanism.
•
Steps 5 and 6 require two people to perform the task.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...