Main Menu data will automatically be saved on the hard disk drive.)
R e c a l i b r a t i o n P r o c e d u r e s
R e c a l i b r a t i o n P r o c e d u r e s
Perform the following recalibration procedures.
#
#
Color Wheel Calibration and Test
1)
Calibrate the color wheel (refer to section 4).
2)
Enter QA Tests. Run the 30/60 mixed tests for 5 minutes. No errors should be reported.
Observe the color wheel for proper operation.
#
Light Intensity Calibrations
The addition of the new color wheel and the blue-yellow hardware requires recalibration of the light
intensities. Refer to the HFA II-i Service Manual as required to complete the calibrations (section
4).
The light intensity calibrations MUST be performed in the order given below. Print both the HFA II-i
White Light and Blue-Yellow Light Verification Data printouts.
1)
White Projector Intensity Calibration
2)
Blue Projector Intensity Calibration
3)
White Bowl Intensity Calibration
4)
Yellow Bowl Intensity Calibration
5)
Wedge Calibration
6)
Blue Correction Calibration
7)
Intensity/Spot Ratio Verification
NOTE:
If the Wedge Calibration is changed after performing the Blue Correction Calibration, an error 130009 will
occur during the power-up sequence. To correct this, perform the Blue Correction Calibration procedure once again.
Again save calibration to the floppy calibration diskette.
O p e r a t i o n a l C h e c k s
O p e r a t i o n a l C h e c k s
To ensure that the upgraded features are accessible to the operator, perform the following
operational checks.
For All Upgrades:
1)
Proceed to a patient test.
2)
Select Change Parameters.
3)
Select Blue-Yellow ON.
4)
Start a 30-2 test. Look into the bowl and ensure that a blue stimulus and yellow
background illumination can be observed.
5)
When the test is complete, save to both floppy and hard drive.
6)
Proceed to File Functions.
7)
Select View Test.
8)
Select Hard Drive.
9)
Recall the 30-2 Blue-Yellow test previously run.
10)
Print out the test results.
I n s t r u m e n t P o s t - C h e c k o u t
I n s t r u m e n t P o s t - C h e c k o u t
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...