16)
Check that the external brightness knob moves freely and has additional
brightness range.
17)
Check that the patient chinrest and headrest move smoothly and completely
up/down and left/right. (Noise will be heard at the end of each motor limit.)
18)
Observe the bowl for dirt or other cosmetically unacceptable spots. Clean as
needed.
19)
Check the operation of the rear fan, and clean the filter
20)
Check that the date is correct and the time is within 2 minutes.
21)
Select a patient test. Change parameters. Verify that the central fixation LED,
inner diamond, and outer diamond all function.
22)
Turn foveal threshold on (except Model 720
i
). Select size III. Start the test.
Verify that the size III spot is in the center of the inner fixation diamond.
Ensure that the spots are sharply focused and no halo is visible. Test the
patient button for operation.
23)
Recall a patient test from your test floppy disk. Display and print the test.
24)
Recall a test from the hard disk. Display and print the test.
___ 25)
Duplicate a patient test from your test floppy disk to a formatted floppy.
Display the copied test.
26)
Copy a patient test from your test floppy disk to the hard drive. Delete that
test from the hard disk.
27)
Perform a backup using the Magneto-Optical drive to verify the backup/restore
procedure.
28)
Turn power off/on and wait for the Main Menu. No errors should be reported.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...