Page 2
FA2i-031A
•
Calibrate all of the spot sizes, as per Section 4.8.2.3 of the Field Service Guide. However, you must
not press the “Store” button after setting each of the positions. They will be held in memory and will
be stored when you press the “Store” button when the spot ratio is in specification.
•
After calibrating the spot sizes, press the “Measured” button.
•
A
“Please Wait”
dialogue will temporarily appear.
•
A new stored value entry then appears below the “Measured” button in the format,
XX.XX
.
•
Check the new stored value. If the new stored value is within specified range (14.1 – 17.9), press
the “Store” button to save the new spot ratio.
•
Press
“
Done”
button to leave the aperture calibration.
Note -
Left Home, Right Home and Offset calibrations will be required after aperture calibration.
No other changes to the calibration process have been created with the HFA II-i revision 4.1 application
software.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...