Page 6
FA2i-032
Additional Networking Troubleshooting Tools/Settings
To locate additional networking troubleshooting tools, select
System Setup -> Additional Setup
-> Diagnostic
. Enter the password 1234 and select Enter. On the
User Diagnostic Menu
select
the
Advanced Network Options
button.
Advanced Network Options
displays the
Enable
startup connection
and the
Set Network Timeout
.
Select
Enable startup connection
for the following features:
Archive/Retrieve Host, Data
Export Host, EMR/PMS Host
and
Worklist Host
.
Archive/Retrieve Host
- Selecting
Archive/Retrieve Host
will activate the system software to
attempt to confirm the connection to the folder on the server setup for the Archive/Retrieve Host
each time the instrument starts up. The practice/engineer can select to have this feature to their
preference.
Data Export Host
- Selecting
Data Export Host
will activate the system software to attempt to
connect to the folder on the server setup for the Data Export Host each time the instrument starts
up. The practice/engineer can select to have this feature to their preference.
EMR/PMS Host
- Selecting
EMR/PMS Host
will activate the system software to attempt to
connect to the folder on the server setup for the EMR/PMS Host
each time the instrument starts
up. The practice/engineer can select to have this feature to their preference.
Worklist Host
- Selecting and
Worklist Host
will activate the system software to attempt to
connect to the folder on the server setup for the and Worklist Host each time the instrument
starts up. The practice/engineer can select to have this feature to their preference.
The
Set Network Timeout
button will bring up a dialogue to select the number of seconds that
the application attempts to connect to the file server before creating a connection error. The
practice/engineer can select to have this feature to their preference.
Networking T roubleshooting Error Messages
Networking Error messages generated on the HFA2i are presented in a tabular format in
appendix F of the HFAII-I User’s Manual. These tables list the specific error messages in one
column and possible solutions in the other. The four error tables are:
Specific Networking
Errors, Unknown Networking Errors, General Networking Errors
and
Worklist Networking
Errors
.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...