Humphrey Field Analyzer II - i
General Information
Confidential and Proprietary
52235C1105
1 - 9
Field Service Guide
5.
If the power cord or plug on the instrument is damaged, a shock or fire hazard may
result. Do not allow continued operation of the instrument until the damaged cord or plug
has been replaced.
6.
To prevent personal injury and damage to the instrument:
P
Use only the power table recommended by Carl Zeiss Meditec.
P
Do not place the instrument on uneven or sloped surfaces.
7.
For stability of the power table, strictly adhere to the following guidelines:
P
Ensure that the instrument is secured to the power table with the screws provided.
P
Do Not
place the instrument near the operator end of the table during setup,
service, or operation.
P
Before servicing the instrument, cycle the table to its lowest position and ensure that
the slider is locked in position.
8.
Ensure that the instrument is installed on a stable, vibration-free surface.
9.
Be cautious when you first touch the projection assembly.
The projection assembly
can become extremely hot
any time the projection bulb is lit for prolonged periods,
such as during sustained patient testing.
10. The bowl lamp voltage is approximately 550 volts — AVOID TOUCHING!
Instrument Precautions
1.
When the instrument is being unpacked, save the shipping materials for possible future
use. Whenever the instrument is shipped, it must be properly packed to prevent damage.
Do Not
place objects in the bowl during shipment of the
HFA II - i
.
2.
When spare parts are received, save the shipping materials for returning the defective
part(s), if appropriate.
3.
Handle interconnecting cables carefully. Many of these are constructed of extremely
small coax cable and are easily damaged.
4.
To avoid possible damage to circuit board components, do not plug/unplug cables while
power is applied to the instrument.
5.
Do Not
use Windex® to clean the touch screen, as it can scratch plastics. The
recommended cleaner is a 50% solution of isopropyl alcohol and water. Otherwise, use
a commercially available anti-static plastic cleaner.
6.
Use extreme care whenever working in or near the bowl to avoid causing marks or
scratches to the inner bowl surface.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...