Humphrey Field Analyzer II - i
Appendix G
Confidential and Proprietary
52235C1105
G - 3
Field Service Guide
G.1 Initializing the Cal/Config Data
This procedure is required whenever a new hard drive has been installed, before calibration
is performed.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu, select
Configuration
. The Configuration Menu
appears.
3)
Select
Initialize the Configuration
. A warning message appears informing
you that all configuration and calibration data will be destroyed.
4)
Select
OK
. A test log screen appears confirming whether or not you want to continue.
Select
Initialize
to erase all contents of the Cal/Config Data. (Select
Quit
to
return to the Configuration Menu.)
5)
“Initialization successful” message will appear.
NOTE:
If you initialize the CAL/Config Data, you must also perform the procedures for
Setting Serial Number
,
Setting the Model/Hardware Options
, and
Setting the Software
Options
described below. Then recalibrate the instrument (see Table 3-1).
G.2 Setting Serial Number
This procedure is required any time the hard drive is replaced or the Cal/Config Data is
initialized. During this procedure you will be instructed to type in the serial number of the
instrument. The serial number can always be found on the label located on the rear cover
assembly.
NOTE:
The instrument’s serial number NEVER changes throughout the life of the
instrument. Instrument serial numbers are critical for tracking instrument reliability,
performance, and customer satisfaction.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu select
Configuration
. The Configuration Menu appears.
3)
Select
Set Serial Number
, then select
Edit
. A ten-key pad appears. Type in the
four-digit model of the instrument. For example, 0750, then select [Enter].
4)
A ten-key pad appears. Type in the six-digit serial number of the instrument. For
example, 001000, then select [Enter].
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...