Appendix C
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
C - 8
52235C1105
• If you chose
Name
:
To select all files on the drive, press
Enter
, and
Select All
. All files on the
disk will be marked with a “
9
” to indicate that they have been selected.
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient’s last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a “
9
” to indicate that they have been selected.
• If you chose Date:
The file directory will be displayed in reverse chronological order. Enter a specific date
(mm-dd-yyyy), then choose
Enter.
The file directory will display tests taken on the
date you selected. Select the test(s) you wish to transfer. All files selected will be marked
with a “
9
” to indicate that they have been selected.
14)
Press
Proceed
. A pop up dialog box will appear advising you of the number of tests
selected to be transferred.
STOP HERE!
Do Not choose
Yes
to start the transfer
process.
On the Receiving
HFA II - i
:
15)
From the Main Menu, select System Setup.
16)
Select the Save/Transmit option.
17)
Select RS-232 Options.
18)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
19)
Press Proceed. The Transmit/Save screen will then appear.
20)
Press the Main Menu icon in the upper right corner of the screen.
21)
From the main menu, select File Functions
22)
Select Transfer.
23)
Select HFA II Serial Cable as the Source.
24)
Choose Hard or Floppy as the Destination, then Proceed.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...