Page 4
FA2i-032
Networking Diagnostics
Access the HFA2i Networking Diagnostics by selecting
System Setup -> Communication
Setup -> Networking Diagnostic
. On the Diagnostic menu you will see the following buttons:
Test Connection, Ping a Host, Loop Back Test, Save Diagnostic Results, Reset
Networking, Show MAC Address, Restore from Archive, Reset Archive
. There is also a drop
down entitled
Remote Host
(Figure 3).
Figure 4. Test Connection Screen.
Test Connection –
The
Test Connection
button initiates a series of tests that will indicate
successful connection to the host, login to the host, access the host folder, read file from the
folder, write to the host directory and delete from the host folder. The success or failure of each
of these tasks can help in diagnosing problems (Figure 4). Each task will be recorded and then
displayed on the setup screen. Prior to selecting the test connection button, ensure that
parameters have been set in the
Communications Setup
screen. Failure to properly setup the
shared folder or FPT site will result in failures during the test connection execution. This is one of
the most powerful diagnostic tools in the diagnostic menu.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...