Adjustment / Calibration
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
4 - 30
52235C1105
10) During the Blue-Yellow Initialization process a percent complete gauge will be dis-
played on screen. The process takes about 4-5 minutes, and when complete, the
message
"Blue-Yellow Initialization Successful"
will be displayed. When the message is
displayed, select
OK
.
11) Select
DONE
.
4.8.4
Print Cal Values
Selecting
Print Cal Values
will print out the current settings stored on the hard drive. (NVM
data), and also the film and glass wedge values. Refer to Appendix H for printout examples
and interpretation instructions.
4.8.5
Miscellaneous
Miscellaneous
is intended for use only by Manufacturing. Selection will display a ten-key
pad. To return to the previous menu, select
Cancel
.
4.8.6
Camera
4.8.6.1
Camera Position / Size
Purpose
This procedure ensures that the video insert on the CRT is the correct size, and at the correct
position.
Camera Position/Size (for all
HFA II - i
models)
Special Tools/Equipment Required
P
Tool Stand
P
Fake Eye
P
Strip of Masking Tape
Prerequisites
1)
Remove the front and rear covers (Sections 3.2 and 3.3); then reinstall the front cover.
2)
Verify the trial lens holder alignment (Section 4.9.6).
3)
Remove the height adaptor from the tool stand and attach the fake eye to the tool
stand.
4)
Mount the tool stand on the chinrest.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...