Adjustment / Calibration
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
4 - 26
52235C1105
3)
After 7 – 10 minutes the unit beeps to indicate the calibration is completed.
Select
STORE
, then
DONE.
C
If the instrument is equipped with the Blue-Yellow option, proceed to Section 4.8.3.4,
Blue Correction.
C
If the instrument is
NOT
equipped the Blue-Yellow option, go to Section 4.8.1 and
verify the after intensity values.
4.8.3.4
Blue Correction
Purpose
This procedure ensures that the wedges are properly calibrated for attenuation of the blue
stimulus light. Blue Correction calibration must be performed whenever blue projector
intensity calibration has been performed or the Wedge calibration has been performed.
During the Blue Correction calibration procedure, the blue filter on the color wheel is
rotated into the light path such that a blue stimulus is reflected into thee detector via the
mirror instead of being projected into the bowl. Each step position on the glass wedge and
the film wedge is measured by the spot detector, compared to the white light calibration
table and a resulting blue stimulus correction table is stored on the hard drive and calibration
floppy diskette.
Special Tools/Equipment Required
P
Tool Stand
P
Mirror Tool
P
Hand-Held Mirror
P
Bowl Shroud
Prerequisites
Before performing Blue Correction, verify the following:
P
Projection lamp filament voltage with lamp on. If voltage is over 10 volts, replace
lamp.
P
Projector Adjustment (Section 4.8.2.1)
P
Shutter Calibration (Section 4.8.2.2)
P
Aperture Calibration (Section 4.8.2.3)
P
Color Wheel Calibration (Section 4.8.2.4)
P
Focus Calibration (Section 4.8.2.7)
P
White Projector Intensity Calibration (Section 4.8.3.1.1) (Required only if the
Projector Lamp position was adjusted.)
P
Blue Projector Intensity Calibration (Section 4.8.3.1.2)
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...