FA2i-012
Page 13
Prior to completing the upgrade, it is essential that you perform an Instrument Post-Checkout. This
will ensure that the instrument is in calibration and all features are operational.
1)
To perform the Instrument Post-Checkout, perform the System Checkout (section 2.1).
S u g g e s t i o n s f o r C u s t o m e r T r a i n i n g
S u g g e s t i o n s f o r C u s t o m e r T r a i n i n g
1)
Ensure that the customer has an understanding of the following instrument operations:
•
How to access Change Parameters and select stimulus color changes.
•
How to access Change Parameters and turn Blue-Yellow to ON.
•
How to run a Blue-Yellow test.
•
How to print out Blue-Yellow Statpac results.
•
How to access the hard disk drive.
•
How to copy files from the floppy disk drive to the hard disk drive.
•
How to select patient files from the hard disk drive.
R e p o r t i n g t h e U p g r a d e t o H u m p h r e y S y s t e m s
R e p o r t i n g t h e U p g r a d e t o H u m p h r e y S y s t e m s
All upgrades performed to the HFA II-i must be reported to Humphrey Systems as described below.
WARNING:
Any unauthorized HFA II-i upgrades (those performed without
notifying Humphrey Systems) will automatically be downgraded to the model
and features that existed prior to the unauthorized upgrade, when a future
revision of software is loaded into the instrument.
1)
Both U.S. Domestic and International (Affiliates and Distributors) must send a fax to
Humphrey Technical Support (925-557-4652) containing the information itemized below
(
Note
!
For your convenience, an appropriate fax form is included at the back of this bulletin)
.
a)
Technician's name and Affiliate/Distributor home office fax number.
b)
The serial number of the HFA II-i that was upgraded.
c)
The customer's name and address.
d)
The type of upgrade performed (see HFA II-i Upgrade Types below).
HFA II-i Upgrade Types:
Only the upgrade covered in this service bulletin is listed below.
740i-to-745i (includes Blue-Yellow) Upgrade
2)
Once Technical Support receives this information, the Order Administration database will be
updated to indicate the new instrument configuration.
If the information is not received,
any subsequent software revision upgrade on the instrument will cancel the
authorized but unreported instrument upgrade, and a second service call will
then be required to restore the instrument upgrade.
Finishing Up
Finishing Up
#
#
Return Paperwork
The following instructions apply to U.S. Domestic Service Representatives. However, it is
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...