27)
Press the
Main Menu
icon to exit this function.
C.5
HFA II - i
To
HFA II - i
Serial Data Transfer
Transfer of serial data from the
HFA II - i
to
HFA II - i
is possible in both the Serial I transfer
mode and Serial II transfer mode. You must have both of the
HFA II - i
units in the Serial II
transfer mode to ensure that all of the test information is passed to the
HFA II - i
.
Transmitting
HFA II - i
Tests to
HFA II - i
1)
Ensure that the Serial II Transfer Disk is loaded on both instruments. Check the
destination window of the Transfer Files selection on the File Functions Menu. If not,
load the Serial II Transfer Disk as described in Appendix section C.2.
2)
Ensure that the data transfer cable is connected between the transmitting
HFA II - i
DATA XFER port and the receiving
HFA II - i
DATA XFER port.
Note:
When the HFA II-i is receiving data, it can distinguish between serial modes
automatically.
On the Transmitting
HFA II - i
3)
From the Main Menu, select
System Setup
.
4)
Select the
Save/Transmit
option.
5)
Select
RS-232 Options
.
6)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
7)
Press
Proceed
. The Transmit/Save screen will then appear.
8)
Press the
Main Menu icon
in the upper right corner of the screen.
9)
From the Main Menu, select
File Functions
.
10)
Select
Transfer Tests
. A dialog box titled “Disk Options” will appear on screen.
11)
Select
Hard Drive
as the Source.
12)
Select
HFA II Serial Cable
as the Destination.
13)
Select the
Directory Order
(either
Name
or
Date
), then press
Proceed
.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...