FA2i-005J
Page 9
KNOWN ERRORS/ ISSUES - Application Software Rev. 3.4.6
•
Blank screens still appear during startup.
•
A database cannot be restored from a backup floppy made on an HFA II. You can only re-initialize a floppy
disk. This is equivalent to a quick format on a PC. If the floppy is unformatted, you will need to format it on
a PC or an HFA2.
•
You can leave a floppy in the drive during boot if you change the CMOS settings so that the instrument first
attempts to boot from the hard drive and then attempts to boot from floppy.
•
Do not remove a database floppy disk while printing from the floppy database.
•
The 3.4.6 application software has the same issues as the C3.1 revision.
SOFTWARE FEATURES - DST Rev. 3.4.6 SSE (P/N 55250)
•
The DST for rev 3.4.6 application software was designed to work with rev 3.4.6 application software only.
It will not be issued to the field. Any request for exceptions should be directed to CZMI Technical Support.
•
No new errors known.
•
Before the GPA software will function, the GPA licensing process must be completed. Details of the licensing
process will be available in the FA2i-025 service bulletin.
•
The “i” screen has been updated to show if the GPA function is enabled or not.
KNOWN ERRORS/ ISSUES - Application Software Rev. 3.5
•
If the date of birth information is entered after the name information in the Patient Information screen and the
date is entered as MM/DD/YYYY, an error display indicating that the DOB is incorrectly input will result. If the
DOB is entered before the patient name or the DOB is entered in the MM/DD/YY format, it will be correctly
displayed.
SOFTWARE FEATURES - Rev. 3.6 - XML Application Software (P/N xxxxx)
[4 - disk set]
•
Not for commercial release — the HFA2P revision 3.6 software provides the application code needed to
export exam data in XML format and the export of print images in bitmap file format.
•
Automatically export the XML files at the end of the exam or at any time on command using the Transfer Tests
menu.
•
Automatically export the XML single page print out in bitmap file format at the end of the exam or at any time
on command using the Transfer Tests menu.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...