FA2i-013A
Page 3
ÚÚÚ
PROPRIETARY and CONFIDENTIAL
ÚÚÚ
TROUBLESHOOTING
Normally rebooting the printer and the HFA2i will restore printing operation. If rebooting does not
restore normal printing, run the self test function. The self test function will often restore normal
printing.
The self test may be run with or without the data cable connected to the HFA2i. To initiate the self
test feature:
1)
Set the printer ON/OFF switch to OFF.
2)
Press in and hold the paper advance button.
3)
Set the printer ON/OFF switch to ON.
4)
Release the paper advance button.
5)
The printer will produce a printout that includes:
a) Model information
b) Firmware version
c) Head test
d) Platen test
e) Character sets
f) Buffer information
g) Analog input
Note –
Be sure to attempt the self test and retry printing prior to replacing a printer.
Self Test Printout
The following page shows a sample self test printout.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...