Appendix H
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
H - 10
52235C1105
HFA II - i
Blue-Yellow Verification Data Printout
(Refer to sample printout on following page)
(13)
Indicates whether the verification data was obtained either BEFORE or AFTER servicing the
HFA II - i
.
(14)
Serial number of the
HFA II - i
.
(15)
Serial number of the Soligor lightmeter being used to perform the verification.
(16)
Revision level of the system software currently installed in the instrument.
(17)
The multiplier values (P3B and P3Y) written on the calibration label and affixed to the
Soligor lightmeter being used to perform the verification. (
Note:
BYP3 and P3 are the older
multiplier values used to calibrate Blue-Yellow intensities. The newer values are P3B and P3Y.
)
(18)
The current blue projector voltage required to obtain a 0 dB light intensity value displayed in
Ft-L.
(19)
The dB spread or amount of dB difference between the highest and lowest light intensity data
points appearing on the graph (including the BOWL). For the Blue Light Verification Data
Printout, this value must be 1.5 dB or less.
(20)
A numerical and graphical representation of each measured spot value (0 dB - 23 dB) and
the BOWL value. The tolerances (MIN and MAX) as well as the actual measured value
(READING) for each data point are indicated in Ft-L.
(21)
The asterisk indicates a value which is out of tolerance. (
Note:
The data points at 18 dB and 23
dB are not used during field service evaluation.
)
(22)
Light intensity data points. Five data points (including the BOWL) are plotted indicating the
attenuation produced by the glass and film wedges.
(23)
Location to be signed by the service engineer performing the verification.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...