Problem
The battery in some HFA II-i computer assemblies may fail prematurely, causing the BIOS settings to be
lost. The instrument may also experience loss of Date and Time settings.
Cause
The CPU battery circuit may discharge the battery faster than normal. This may prevent the battery from
maintaining the proper charge, resulting in the above stated problems.
Solution
The CPU vendor has improved the battery circuit. We have also implemented new testing procedures on
the production line.
If the instrument loses the BIOS settings, or is losing date and time, the voltage of the battery should be
tested. If the battery voltage is less than 3.0 volts, then the CPU/Backplane enclosure should be replaced.
Note:
If the battery voltage is less than 3.0 Volts, and the battery has been in the system for three or
more years, you should replace only the battery.
The serial number ranges, of instruments that may be affected, are listed below. Not all instruments within
the serial number range will be affected.
720i-5020 to 720i-5061
740i-7020 to 740i-7204
750i-8011 to 750i-8171
The CPU/Backplane enclosure should only be replaced on failure. Credit will only be given for failed
enclosures, which have been returned to CZMI for inspection. The entire CPU/Backplane Enclosure must
be returned (see service bulletin FA2i-011).
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...