Page 2
FA2i-015
For further clarification, please contact Customer Service toll free at 877-486-7473. We
appreciate your assistance.
Sincerely,
Jeff Durban
Product Manager
Enclosures
Item 1:
“File/Name” Issue
If an operator fails to follow the user instructions, and under a highly specific set of unusual
circumstances (further described below), then it is possible to have an incorrect patient name
associated with a visual field test performed on the Humphrey Field Analyzer II -
i
series with
system software revisions C2, C3, C3_MO or C3.1. For example, Patient A’s name may be
matched with Patient B’s test if
all
of the following circumstances are met:
If, during the course of testing a patient on your HFA II-
i
series:
•
the instrument encounters an error where the instrument did not restart,
•
and
the patient being tested is not the first patient of the day to be tested, (ie, Patient B)
•
and
you respond to the error message by selecting the “OK” button,
•
and
you then select the “Test Other Eye” button,
then the instrument will revert to displaying the name of the last patient who completed a visual
field examination on the instrument (ie, Patient A).
If, after the above conditions occur, you do not observe the incorrect patient name displayed on
the screen:
•
and
you elect to resume the examination,
•
and
the test is completed successfully,
•
and
you elect to save the test result,
•
and
you select “Yes” in response to the message presented on the screen asking you to
confirm that the patient’s name and date of birth are correct [“Do You Wish To Save This
Test For”: (ie, “Patient A”)], failing to observe the incorrect patient name displayed,
then the file saved will have the incorrect patient’s name recorded. Additionally, if you chose to
print the visual field test results, the incorrect patient name would be printed on the visual field
test results.
Action Required: Install system software revision 3.2;
check and correct any suspect files.
If you should have any files where you suspect that the patient’s visual field result is not properly
associated with the patient’s name, and if, after further investigation you determine that the file
must be updated, you may correct the file by completing the following steps:
Main Menu>File Functions>Change Patient Data>One Test>Select the Source (Hard Drive,
Floppy Drive)>Select Directory Order (Name, Date or Patient)> Proceed>Enter Name to
Find>Enter>Select the patient from the file directory>Proceed>Edit Patient
Information>Select Enter after each field is updated>Select Proceed. To exit the Change
Patient Data screen, select the Main Menu icon.
Note: You may wish to reprint the visual field test results with the correct patient information
recorded. Please be aware that merely installing the new system software revision 3.2 will
not
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...