FA2i-005J
Page 11
SERVICE FEATURES - Application Software Rev. 4.0
•
Licensing of Network, NetPro, GPA and SITA SWAP.
•
Configuration of Instrument for Network and NetPro.
•
Database Upgrade.
•
Save to floppy the text files of system log, information screen, and calibration values.
•
New tools for evaluating Networking parameters.
SERVICE CONCERNS - Application Software Rev. 4.0
•
Data created on a HFA 2i with a revision 4.0 software will not be able to be saved to a HFA2i with
software revision lower than 4.0 or to any HFA II system.
See page 2 of the HFA II-i System Software
Upgrade document P/N 58122.
•
Database upgrade may not be possible on existing hard drive partition due to the partition being nearly full.
If the existing 2.0 gigabyte partition is nearly full, a second partition may be created on the hard drive or
replacement of the hard drive with a drive that can be repartitioned to meet the system needs. A separate
service bulletin will cover this topic.
KNOWN ERRORS/ ISSUES - Application Software Rev. 4.0
•
None.
SOFTWARE FEATURES - Rev. 4.1 Application Software (P/N 60138)
Changes
•
Revision 4.1 software was designed to be compatible with the new EMI Box and the new Motor Driver Board.
If either the new EMI box (P/N 60673) or the new MDB are used, the 4.1 software must be loaded. Details
are presented in HFA2i-036 and HFA2i-038.
•
Changes to Touch Screen Calibration – Rev 4.1 will beep when you have held your finger on the calibration
target for the required amount of time.
•
Startup Test to Determine Calibration of Aperture – At startup of the instrument, an evaluation of correct
aperture positioning takes place. Failure of the test will result in a Startup State Error 19 error that says
Aperture may be misaligned. Pressing Proceed will not allow testing. Recalibration of the aperture and
measurement of the spot ratio with a value within specification, will be required before testing can proceed.
•
Startup Test to Determine Shutter Misalignment – At startup of the instrument, an evaluation of correct shutter
positioning takes place. Failure of the test will result in a Startup State Error 18 error that says Shutter may
be misaligned. Pressing Proceed will not allow testing. Recalibration will be required before testing can
proceed.
New Features
•
There are no new features or enhancements in the HFA2i Rev 4.1 software.
Fixes
•
The failure to save to a floppy disk at the end of the test when the system gives the “floppy full, please insert
a new formatted floppy” message, has been corrected. With the rev 4.1 software the tests will be saved when
the new formatted floppy is inserted after the prompt.
•
Long patient names are now displayed in accordance with the industry standards.
•
The setting of the default single field printout style for instruments with GPA will default to GPA rather than
classic printout once the GPA feature is licensed.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...