Adjustment / Calibration
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
4 - 10
52235C1105
White Bowl Intensity Verification:
12) Select
#61/BOWL
. The bowl lamp warm-up period will begin. (You may cancel from
the warmup period when the voltage stabilizes at >20%.)
After the warmup period ends or when the voltage reading has stabilized:
13) Enter the voltage reading via the ten-key pad.
14) If the
HFA II - i
is equipped with the Blue-Yellow option, proceed to the Blue Spot
Intensity Verification below.
If the
HFA II - i
is
NOT
equipped with Blue-Yellow option, select
to print out the
HFA II - i
White Light Verification Data printout. Then proceed to Cal/Wedge printout
below.
Blue Spot Intensity Verification:
15) Select
BLUE/YELLOW
, and then
#62/0dB
.
16) Ensure that the light meter is set to
position 3.
17) Position the light meter on the projected spot, then maximize the reading by position-
ing the chinrest via the chinrest control switch. (First left/right, then up/down, then
refine with left/right again. If you subsequently move the light meter accidently, repeat
this step in order to remaximize the meter reading.)
18) Enter the #62/0dB voltage reading via the ten-key pad on the screen and select
[Enter]. The calculated foot-lamberts will be displayed. Also displayed is the tolerance
for each reading. If an asterisk (*) appears, this indicates a reading that is out of
tolerance.
19) Repeat the procedure for each of the selections #63/3dB thru #65/13dB. (#66 and
#67 are not used at this time.)
Yellow Bowl Intensity Verification:
20) Select
BLUE/YELLOW
and select
BOWL
.
21) Allow 3 minutes for lamp warm-up and ensure that the light meter is set to
position
3.
22) Enter the BOWL voltage reading via the ten-key pad on the screen.
23) Select
DONE
and then
PRINT BOTH
to print out both the
HFA II - i
White Light and
Blue-Yellow Light Verification Data printouts. Then proceed to Cal/Wedge Printout
below.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...