Humphrey Field Analyzer II - i
General Information
Confidential and Proprietary
52235C1105
1 - 5
Field Service Guide
1.3
HFA II - i
Service Strategy
1.3.1
Two-Level Service Strategy
A two-level service strategy is used for the
HFA II - i
: Level 1 for on-site service; and Level 2
for Repair Center service. Level 1 service employs modular replacement, wherein faulty
circuit boards and certain other assemblies are replaced rather than repaired on-site. These
faulty assemblies are shipped to a Carl Zeiss Meditec Repair Center for repairs. There are
also certain procedures that require special equipment available only at a Repair Center.
Designated Repair Centers (currently Dublin, CA and Jena, Germany) are the second level
of service for the
HFA II - i
. The Repair Centers perform major circuit board troubleshooting
and repair, plus any other service action that requires special equipment or procedures not
available in the field.
Several of the circuit boards in the
HFA II - i
are multilayer boards and use Surface Mount
Technology (SMT) components. These boards require special equipment and techniques for
troubleshooting and repair.
All service procedures (including instrument calibration) can be performed in the field, except
for those listed below. For Carl Zeiss Meditec U.S. Domestic operations, the following
procedures must be performed at the Carl Zeiss Meditec Repair Center.
P
Circuit board troubleshooting and component replacement
P
Repair of floppy, tape, and hard drives.
P
Alignment of projection carriage rails and first projection mirror (top turret mirror)
P
Repair of power supply assembly
P
Repair of camera assembly
1.3.2
Three Steps to Completing an
HFA II - i
Service Call
The basic approach to an
HFA II - i
service call is outlined below. This typical process
includes collection of general instrument calibration data and light intensity data both
Before
service and again
After
service. The process is described in detail in Section 4.8.1. For
guidelines, refer to
HFA II - i
Field Service Paperwork Requirements (Section 1.3.3).
1.
Obtain the
Before
Light Intensity instrument data (4.8.1).
This step assumes that the
HFA II - i
is operable; that is, it will power up to the
Main Menu without error. This data gives the service representative a base from
which to evaluate the light intensity operation of the instrument, and a point of
comparison if recalibration is required.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...