General Information
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
1 - 22
52235C1105
C
Uninterruptible Power Supply (UPS)
Carl Zeiss Meditec does not sell a UPS (provides battery backup during power loss). However,
when selecting a UPS for use with the
HFA II - i
, note the following specifications and
recommendations:
C
The UPS must have a minimum 450 VA rating.
C
It must be dedicated to the
HFA II - i
. Do not connect the power table or other
devices into the UPS.
C
Recommended model: APC smart UPS 620.
Power backup during power loss will depend on the type of UPS selected. Consult a local
supplier for details.
C
HP LaserJet Printer
Carl Zeiss Meditec does not sell HP LaserJet Printers. Refer to Appendix D, Approved Parallel
Printers for details on using the HP LaserJet.
1.9
Specifications
HFA II - i
Field Analyzer
Physical Characteristics
Dimensions . . . . . . . . . . . . . . . . . . 23" H x 22.25" W x 19.25" D
(58.4 x 56.5 x 48.9 cm)
Weight . . . . . . . . . . . . . . . . . . . . . 87.6 lbs (36.3 kg)
Recommended Room Size . . . . . . 106" x 64" (269.2 x 162.6 cm)
(Ventilation recommended)
Environmental
Operating Temp (ambient) . . . . . . 41 to 95
E
F
Humidity (ambient) . . . . . . . . . . . . 20 – 85%
Heat Generated . . . . . . . . . . . . . . 1,570 BTU/Hr (460 W)
Electrical
Input Power . . . . . . . . . . . . . . . . . 100 – 120 V @ 4 A, 50-60 Hz
230 V @ 2 A, 50 Hz
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...