Troubleshooting
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
5 - 4
52235C1105
#
Solid Failure
- the problem presents itself continuously.
#
Intermittent Duplicable
- the problem is intermittent in nature. Usually a specific
sequence of events reproduces the intermittent symptom, although occasionally
these problems may be random in nature. By performing the proper sequence of
events, the problem can usually be reproduced. If a sequence of events readily
reproduces the problem, examination of this sequence of events may help identify
the faulty assembly.
#
Intermittent Non-Duplicable
- the problem is intermittent, and all reasonable
attempts at recreating the symptoms fail. These are the most difficult problems to
resolve.
3)
Check for Field Service Bulletins that Address the Problem.
If a service bulletin addresses a resolution, be sure to take the necessary equipment to
perform the changes.
While On-Site:
4)
Have the Operator Demonstrate the Problem.
This is very important. Before doing anything, always have the operator attempt to
demonstrate the problem to you. By having the operator demonstrate the problem,
you may solve the problem immediately, or avoid some incorrect assumptions. Having
the operator demonstrate the problem will
ALWAYS
result in one of the following:
#
Reveal Operator Error
- You may immediately see that the cause of the problem
is operator error. You then have the opportunity to train the customer in the
proper use of the instrument, instead of spending time troubleshooting a
nonexistent instrument problem.
#
Demonstrate specific problem sequence
- The operator will show you the
specific sequence of events needed to reproduce the problem, reducing your
diagnostic time.
#
Help Form Ideas
- The operator may not be able to duplicate the problem.
However, the sequence of steps used to demonstrate the problem will give you
ideas on how to pursue the problem in other ways. In this case you have the
opportunity to explain to the customer the difficulty in resolving a nonduplicable
problem.
5)
Resolve the Obvious.
Obvious mechanical or physical defects should be resolved when discovered. Often,
resolving the obvious will lead you to the real problem.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...