Page 12
FA2i-005J
ÚÚÚ
NOTICE
ÚÚÚ
Any unauthorized HFA II upgrades (those performed without notifying Carl
Zeiss Meditec Inc.) will automatically be downgraded to the model and
features that existed prior to the unauthorized upgrade, when a future revision
of software is loaded onto the instrument.
•
The C10-2 tests will now appear on the printout if a SFA printout is selected when the 10-2 test was the most
recent.
•
The instrument will no longer crash when an invalid date is encountered in a worklist. The date will be
discarded.
•
The system will not crash when it encounters an option returned from the DHCP server that is not supported
by RFR2132 standard. Unsupported options will be deleted in rev 4.1.
SERVICE FEATURES - Application Software Rev. 4.1
•
Licensing of Network, NetPro, GPA and SITA SWAP will be required to enable these features.
SERVICE CONCERNS - Application Software Rev. 4.1
•
Startup State Errors 18 and 19 may be generated on first power up of the instrument after loading the rev 4.1
software. See Changes above for details.
•
In the Mechanical Calibration Menu, a new section titled V/III Spot Ratio, has been added. This section
displays the currently stored value and has a button to measure the spot ratio for spot sizes V (5) and III (3).
A service bulletin will be released to address the new calibration procedure.
•
Data created on a HFA 2i with a revision 4.0 software will not be able to be saved to a HFA2i with software
revision lower than 4.0 or to any HFA II system. See the HFA II-i System Software Upgrade document.
•
If the existing 2.0 gigabyte partition is nearly full, a second partition may be created on the hard drive or
replacement of the hard drive, with a drive that can be re-partitioned to meet the system needs. Service
bulletin FA2i-037 will cover this topic.
SOFTWARE FEATURES - Diagnostic Service Tool Disk Rev. 4.1.1 (P/N 63025)
Changes
•
DST Revision 4.1.1 software was designed to be compatible with the new EMI Box, BIOS, and the new Motor
Driver Board that are associated with the release of revision 4.1 software. DST rev 4.1.1 will work with all
HFA2i hardware and software configurations. There are no new features for the DST 4.1.1 revision.
SOFTWARE FEATURES - BIOS Disk SBEP 110J (P/N 61753)
Changes
•
The SBEP110J BIOS disk was designed to support the new backplane. The SBEP110J BIOS is compatible
with the new and old backplanes, EMI Boxes and Motor Driver Boards.
•
Under most circumstances you will not need to load the SBEP110J BIOS. The SBEP110J BIOS will be loaded
on all new EMI boxes with part number 60673. The only circumstance that would require loading the
SBEP110J BIOS would be a suspicion of corruption to the existing BIOS.
Содержание humphrey HFA II-i series
Страница 4: ......
Страница 152: ...Troubleshooting Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 5 2 52235C1105 Notes...
Страница 178: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 2 52235C1105 Notes...
Страница 179: ......
Страница 180: ...Diagrams Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 6 4 52235C1105...
Страница 192: ...Parts Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide 7 2 52235C1105 Notes...
Страница 207: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 17 Field Service Guide...
Страница 221: ...Humphrey Field Analyzer II i Parts Confidential and Proprietary 52235C1105 7 31 Field Service Guide 7 7 9 Bowl Assembly...
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 244: ...Appendices Humphrey Field Analyzer II i Confidential and Proprietary Field Service Guide App 4 52235C1105 Notes...
Страница 313: ...Humphrey Field Analyzer II i Appendix J Confidential and Proprietary 52235C1105 J 3 Field Service Guide...
Страница 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Страница 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Страница 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Страница 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...