Table top
38
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
12.1
Table top
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Remove the pad plate [1] from the seat section (see
chapter 12.2 on page 40). Important: keep the back section
pad plate attached so that the bar adjustment is retained.
4. Remove the cover [2] from the center box (see chapter 12.3
on page 41).
5. Be sure to note the cable routing.
Disconnect all plugs from the motor controller circuit board [3]
and remove the circuit board [3] from the center box [6] (3
screws). Note the cable adapter for the ISM module
2)
. Lift the
circuit board slightly and pull the cable adapter plug from the
underside of the circuit board.
6. Be sure to note the cable routing.
Disconnect all plugs from the communication processor circuit
board [7] and remove the circuit board [7] from the center
box [6] (4 screws).
7. Carefully push cables W164, W173 and sensor cable
downward out of the center box.
Release the table top [4] from the column: Remove 10
screws [5] from the center box [6].
9. Workplace safety: Due to the heavy weight of the table top,
take care not to strain your back when lifting and putting it
down! A second person is required for lifting and putting down
the table top.
Carefully lift the table top off the column and place it down on
a flat, soft surface.
Assembly
1. Workplace safety: Due to the heavy weight of the table top,
take care not to strain your back when lifting it! A second
person is required for lifting and installing the table top.
Place the table top on the column can attach the center box to
the mounting plate of the column (10 screws, note the different
lengths). When placing the table top, orient it to the openings
2)
not available on all operating table versions
8.
CAUTION
Risk of injury!
Do not lean on the table top if the screws have been removed.
The table top can fall from the column due to the load. Have
a second person support the table top.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......