Toothed belt from the main drive
102
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
toothed belt tension with the tensioning screw (hexagonal
screw).
b) Tighten the screws on the lift motor mounting bracket with 7
Nm (4 screws).
c) Check the toothed belt tension and reset if there are
differences. The value on the belt tension meter must be
140 Hz (±5 Hz).
8. Attach the subfloor cover (4 screws).
9. Safety note: The operating table is extremely heavy; take extra
care not to strain your back while righting it. A second person
is needed to right the operating table. Clear the floor in front
of the operating table and clean it if necessary. Right the
operating table with the help of a second person.
10. Remove the multifunction tool.
11. Remove the toothed wheel.
12. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
13. Lock operating table (brake)
14. Put on the pad.
15. Calibrate the lift motor (CAN test center). See chapter 24.7 on
page 268.
16. Update the software (see the software description from the
CAN test center).
TEST
Perform function test and final check.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......