Cable W164
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
209
21.5
Cable W164
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Remove the table top (see chapter 12.1 page 38).
4. Remove the bellows (see chapter 14.3 on page 78).
5. Grasp the column cover, remove the holding screw from the
Trendelenburg assembly and guide the cover carefully
downward.
6. Open the column cover at the top and slide it down (see
7. Remove the metal panel with the column keypad (see
chapter 14.2 on page 77).
8. Remove the frame (connection ring) (see chapter 14.4 on
9. Unscrew the speaker [1] from the suspension plate [2] (from
above, 2 screws [3]) and place them securely on the column.
10. Note the cable routing and attachment.
Unfasten cable W164 [4] under the mounting plate [5].
11. Note the cable routing and attachment.
Unfasten cable W164 [4] from the suspension plate [6].
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......