Cable W104
200
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
21.1
Cable W104
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Detach the bellows from the cover, carefully guide it up and
affix it (see chapter 9 on page 31).
4. Grasp the column cover, remove the holding screw from the
Trendelenburg assembly and guide the cover carefully
downward.
5. Open the column cover at the top and slide it down (see
6. Remove the metal panel with the column keypad (see
7. Separate the frame pieces [1]/[2] from one another (2
screws [3] from below).
8. Detach frame piece 2 [2] from the tilt left assembly [5] (2
screws [4] with washers).
9. Remove the cable deflector [6] from the power supply unit [7].
To do this, unscrew the 2 screws [9] at the top of the power
supply unit [7] and loosen the two screws [8] on the column
base. Guide the cable deflector out of the screws [8] on the
column base and remove it.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......