Trendelenburg assembly
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
119
3. Remove the table top (see chapter 12.1 on page 38).
4. Remove the bellows (see chapter 14.3 on page 78).
5. Grasp the column cover, remove the holding screw from the
Trendelenburg assembly and guide the cover carefully
downward.
6. Open the column cover at the top and slide it down (see
7. Remove the metal panel with the column keypad (see
chapter 14.2 on page 77).
8. Remove the frame (connection ring) (see chapter 14.4 on
9. Disconnect plug and shielded connection of cable W110
from the Trendelenburg motor.
10. Be sure to note the cable routing.
Remove the cable attachment from the Trendelenburg
assembly and the suspension plate.
11. Unscrew the column controller [1] from the tilt right
assembly [2] (1 screw with washer), carefully disconnect it
from the plug contact (W164) on the Trendelenburg
assembly [3] and place on non-conductive material of the
column.
12. Be sure to note the cable routing.
Unscrew the holder with plug contact (W164) [4] from the
Trendelenburg assembly [3] (3 screws with washers).
13. Be sure to note the cable routing.
Remove the cable attachment for the sensor cable under the
mounting plate.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......