Bleeding the hydraulic system
162
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
button because there is still too little pressure in the system. The
system is bled when the floor locks extend fully (25 mm / 35
mm). The air is pulled into the oil container and maximum
pressure builds during locking.
Make sure the floor locks retract after the procedure
(necessary to right the table).
9. Install the two upper locking cylinders.
10. Restore the original cable routing. Check all connections on
the jack unit to be sure they are secure.
Carefully insert the cover with the assembly into the table base
and mount it (4 screws with washers).
11. Safety note: The operating table is extremely heavy; take extra
care not to strain your back while righting it. A second person
is needed to right the operating table. Clear the floor in front
of the operating table and clean it if necessary. Right the
operating table with the help of a second person.
12. Replace the label over the
E
MERGENCY
RELEASE
button.
13. Remove the multifunction tool.
14. Remove the toothed wheel.
15. Lock operating table (brake)
16. Put on the pad.
TEST
Perform function test and final check.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......