Cable W134
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
213
21.6
Cable W134
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Remove the cable deflector [1] from the power supply unit [2].
To do this, unscrew the 2 screws [4] at the top of the power
supply unit and loosen the two screws [3] on the column base.
Guide the cable deflector [1] out of the screws [3] on the
column base and remove it.
4. Pull both plugs of cable W134 [5] from the power board
(PB1) [6] on the power supply unit.
5. Note the cable routing and attachment.
Unfasten cable W134 [5] from battery 1 [7] up to the
connector board [6] on the power supply unit and remove
cable W134.
Assembly
1. Restore the original cable routing.
Route cable W134 in the column pot.
2. Insert the plugs (2x) of cable W134 into the sockets on the
power board (PB1) (power supply unit) (see circuit diagram
3. Note the shield support.
Refasten cable W134 from the power supply unit up to battery
1.
4. Set the cable deflector onto the two screws on the column
base and slide it flush against the power supply unit. Mount the
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......