Power input fuses
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
155
17.4
Power input fuses
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Lift the cover [1] of the line power socket on the table base.
Pull out the fuse chamber [2], open it, and remove the fuse [3].
4. Check each fuse and replace if necessary.
Assembly
1. Lift the cover of the line power socket on the table base. Pull
out the fuse chamber, open it, and insert the checked fuses (see
2. Close and reinsert the fuse chamber.
3. Close the cover on the line power socket.
4. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
5. Put on the pad.
TEST
Perform function test and final check.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......